Imihlahlandlela yokuvuselela ama-hemodialyer

Ngenxa yezingozi ezingaba khona ezibandakanyekayo ekuphindisweni, okungase zibeke izingozi zokuphepha ezigulini, kunemithetho esebenzayo yokusebenza ngokusebenzisa kabusha ama-hemodialyzers egazi. Abasebenza ngokuqeqeshwa kufanele baziqeqeshe kahle futhi banamathele kwimihlahlandlela yokusebenza ngesikhathi sokuphindisela.

Uhlelo lokwelashwa kwamanzi

Reprocessing must use reverse osmosis water, which must meet biological standards for water quality and meet the water demand of equipment working during peak operation. Ubungako bokungcola okubangelwa amagciwane kanye nama-endotoxins emanzini e-RO kufanele kuhlolwe njalo. Ukuhlolwa kwamanzi kufanele kwenziwe lapho noma eduze kokuhlangana phakathi kwe-dialyzer yegazi kanye nohlelo lokuphinda luvuswe kabusha. Izinga lamagciwane alikwazi ukuba ngaphezulu kwama-200 CFU / ML, ngomkhawulo wokungenelela kwe-50 CFU / ML; Izinga le-Endotoxin alikwazi ukuba ngaphezulu kwe-2 EU / ML, ngomkhawulo wokungenelela kwe-1 EU / ML. Lapho umkhawulo wokungenelela ufinyelelwa, ukusetshenziswa okuqhubekayo kohlelo lokwelashwa kwamanzi kwamukeleka. Kodwa-ke, kufanele kuthathwe izinyathelo (njengokukhipha amagciwane ohlelweni lokwelashwa kwamanzi) ukuvikela ukungcoliswa okwengeziwe. Ukuhlolwa kwe-bacteriological kanye ne-endotoxin yekhwalithi yamanzi kufanele kwenziwe kanye ngesonto, futhi ngemuva kokuhlolwa kwezimbili okulandelanayo kuhlangabezana nezidingo, ukuhlolwa kwe-bacteriological kufanele kwenziwe njalo ngenyanga, futhi ukuhlolwa kwe-endotoxin kufanele kwenziwe okungenani kanye izinyanga ezintathu.

Uhlelo lokuphinda luvundisiwe

Umshini wokuphindaphinda kufanele uqinisekise ukuthi le misebenzi elandelayo: Ukubeka i-dialyzer esimweni esiphindaphindayo se-ultrafiltration sombuso ophindaphindayo wegumbi legazi ne-dialysate Chamber; Ukwenza izivivinyo zobuqotho bokusebenza kanye noMembrane ku-dialyzer; Ukuhlanza igumbi legazi ne-dialysate Chamber ngesisombululo sokubulala amagciwane okungenani amahlandla ama-3 wevolumu yegazi, bese ugcwalisa i-dialyzer ngesisombululo esisebenzayo sokufakelwa phansi kwegciwane.

Wesley's dialyzer reprocessing machine--mode W-F168-A/B is the first full-automatic dialyzer reprocessing machine in the world, with automatic rinse, clean, test, and affuse programs, which can complete dialyzer flushing, dialyzer disinfection, testing, and infusion in about 12 minutes, fully meeting the standards of reuse dialyzer processing, and print the TCV(Total Cell Volume) test result out. Umshini we-dialyzer othomathikili othomathikhi wenza lula umsebenzi wabasebenza futhi uqinisekisa ukuphepha nokusebenza kahle kwama-dialyzer asetshenzisiwe.

W-f168-b

Ukuvikelwa komuntu siqu

Wonke umsebenzi ongathinta igazi leziguli kufanele athathe izinyathelo zokuphepha. Ku-dialyzer Repacessing, opharetha kufanele bagqoke amagilavu ​​nezingubo zokuvikela futhi bathobele amazinga okuvimbela ukutheleleka. Lapho wenza inqubo yokuphuza ubuthi noma isixazululo esidabukisayo, abahlinzeki kufanele bagqoke imaski nabaphenyiso.

Isidingo samadayidi egazi aphindisele

Ngemuva kwe-dialysis, i-dialyzer kufanele isuswe endaweni ehlanzekile futhi iphathwe ngokushesha. Uma kwenzeka kunezimo ezikhethekile, ama-hemodialyzers egazi angalashwa ngamahora ama-2 angaqashelwa ngemuva kokuvuthwa, kanye nezinqubo zokubulala inzalo kanye nenqubo yokubulala inzalo yegazi.

●Rinsing and cleaning: Use standard RO water to rinse and clean the blood and dialysate chamber of the blood hemodialyzer, including back-flushing. I-hydrogen peroxide, i-sodium hypochlorite, i-acid ye-peracetic, namanye ama-reagents amakhemikhali angasetshenziswa njengama-ejenti wokuhlanza i-dialyzer. Kepha, ngaphambi kokungeza amakhemikhali, amakhemikhali angaphambili kufanele asuswe. I-Sodium Hypochlorite kufanele isuswe kwikhambi lokuhlanza ngaphambi kokungeza ngokusemthethweni futhi lingahlanganiswe ne-peracetic acid.

● Ukuhlolwa kwe-TCV kwe-Dilyzer: I-TCV ye-Dialyzer yegazi kufanele ibe nkulu noma ilingane ne-80% ye-TCV yasekuqaleni ngemuva kokuphindisela.

● I-Direyzer Disninfection kanye ne-inzalo: I-hemodialialzer yegazi ehlanziwe kumele ibulawe amagciwane ukuvikela ukungcoliswa kwe-microbial. Both the blood chamber and dialysate chamber must be sterile or in a highly disinfected state, and the dialyzer should be filled with disinfectant solution, with the concentration reaching at least 90% of the regulation. The blood inlet and outlet and the dialysate inlet and outlet of the dialyzer should be disinfected and then covered with new or disinfected caps.

 

● Isitoreji: Ama-diallyzers acutshunguliwe kufanele agcinwe endaweni ebekiwe ukuze ahlukane ne-dialyzers engafakwanga uma kwenzeka ekungcolisweni nasekusetshenzisweni kabi.

Ukubukeka kwangaphandle kuhlola ngemuva kokuphindisela

(1) Alikho igazi noma elinye ibala ngaphandle

(2) Akukho cranny egobolondweni kanye nembobo yegazi noma i-dialysate

(3) Ayikho i-tlotting ne-fiber emnyama ebusweni befayifa eliyize

(4) Akukho ukugcwala ama-terminals amabili e-dialyzer fiber

(5) Thatha ama-cap on the inlet kanye nokuphuma kwegazi ne-dialysate futhi uqiniseke ukuthi akukho ukuvuza komoya.

.

Ukulungiselela phambi kwe-dialysis elandelayo

● Flush i-distinfectant: I-dialyzer kumele igcwaliswe futhi idonswe ngokwanele ngosawoti ojwayelekile ngaphambi kokusetshenziswa.